IAQ Applications |
CT Water Treatment |
Odor & Toxic Gas Removal
Process Controllers |
Calibration Gases
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Ozone Generators |
VOC Monitoring |
Ozone Monitors |
CO2 Monitoring |
Aspiration Devices
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
IAQ Applications : Ozone Monitors
MODEL C-30ZX
FOR HVAC AND INDOOR AIR QUALITY SYSTEMS, WATER TREATMENT PLANTS, PULP BLEACHING MILLS, OZONE GENERATOR MONITORS, PHOTOCOPIER AND LASER PRINTER CENTERS, FUMIGATION PROJECTS, FIRE RESTORATION PROJECTS, VEHICULAR POLLUTION MONITORS, RESEARCH LABS, PILOT PLANTS AND WHERE EVER OZONE EXPOSURE IS POSSIBLE
The C-30ZX has very fast and continuous Response. It is rugged and versatile. There are no Touchy controls.
BENEFITS
Constantly monitors the work environment
Visual indication of ozone concentration by a multi color graphical bar display.
Audible and visual alarm setting capability.Easily understood by non-technical personnel.
Low maintenance.
FEATURES
LED readout changes color as ozone level changes.
Audio alarm and out put for data logger and BMS
Connection to external equipment control
ACCESSORIES
Data loggers
Calibrators
Protective enclosures
Duct mounting kits
AC adaptors
SPECIFICATIONS
Range: 0.02 - 0.14 ppm of ozone (LED scale); 0.02 - 30 ppm via external data logger.
Bar graph Display: Normal (green), 0.05 ppm (Yellow) 0.10 ppm (Red)
Response time: within 10's of seconds of ozone reaching the sensor
Measurement principle: HMOS (heated metal oxide semiconductor) sensor
Size: 31/4" x 13/8" x 21/8" (85 x 35 x 60 mm)
Weight: 5 oz (140 grams)
Power requirements: 12-24 volts DC at 300mA.
Outputs: LED bar graph, audio alarm, 0-3 volt analog output for data loggers and BMS, relay output NO/NC. Alarm actuation and relay operation is programmable
|
|
|